Title Classification of Normal Subjects and Pulmonary Function Disease Patients using Tracheal Respiratory Sound Detection System
Authors 임재중(Im, Jae-Jung) ; 이영주(Lee, Yeong-Ju) ; 전영주(Jeon, Yeong-Ju)
Page pp.220-224
ISSN 1975-8359
Keywords respiratory sound ; trachea ; pulmonary disease ; pulmonary function test
Abstract A new auscultation system for the detection of breath sound form trachea was developed in house. Small size microphone(panasonic pin microphone) was encapsuled in a housing for resonant effect, and hardware for the sound detection was fabricated. Pulmonary function test results were compared with the parameters extracted from frequency spectrum of breath sound obtained from the developed system. Results showed that the peak frequency and relative ratio of integral values between low(80∼400Hz) and high(400∼800Hz) frequency ranges revealed the significant differences. Developed system could be used for distinguishing normal subject and the patients who have pulmonary disease.