• 대한전기학회
Mobile QR Code QR CODE : The Transactions of the Korean Institute of Electrical Engineers
  • COPE
  • kcse
  • 한국과학기술단체총연합회
  • 한국학술지인용색인
  • Scopus
  • crossref
  • orcid
Title A Study on Improvement of MUAP Resolution using Spatial Filter
Authors 양덕진(Yang, Duck-Jin) ; 전창익(Jun, Chang-Ik) ; 이영석(Lee, Young-Suk) ; 이진(Lee, Jin) ; 김성환(Kim, Sung-Hwan)
Page pp.55-64
ISSN 1975-8359
Keywords electromyography(EMG) ; spatial filter ; MUAP
Abstract Conventional bipolar surface electromyography(EMG) technique detects only the superimposed electromyographic activity of a large number of motor units due to its low spatial resolution. For the diagnosis of neuromuscular disorder, the information of single MU is required. In this paper, 9 channel array surface electrode system was as designed and MLoG filter was proposed. Also the MCPT(modified convolution processing technique)method was proposed for the improvement of MUAP resolution. For performance evaluation, power spectrum analysis of random data and raw EMG signal comparison of MUAP shape and quantitative estimation of SNR were executed. As a result, the MUAP resolution improvement of 32% was obtained from the standpoint of the signal-to-noise ratio(SNR).