• 대한전기학회
Mobile QR Code QR CODE : The Transactions of the Korean Institute of Electrical Engineers
  • COPE
  • kcse
  • 한국과학기술단체총연합회
  • 한국학술지인용색인
  • Scopus
  • crossref
  • orcid
Title Estimation of Non- Invasive Blood Pressure Using Peripheral Plethysmograph
Authors 정인철(Jeong In-cheol) ; 신태민(Shin Tae-min) ; 윤형로(Yoon Hyung-Ro)
Page pp.504-509
ISSN 1975-8359
Keywords Blood Pressure ; Pulse Transit Time ; Reflected Wave Arrival Time ; Volume Pulse ; Regression Equation
Abstract This paper presents a new method for obtaining the noninvasive and unrestrained blood pressure readings noninvasively and unrestrainedly using based on reflected wave arrival time(RAT) in the volume of pulse. Since this new method employs only volume pulse, is more rapider and simpler than the method using pulse transit time(PTT) because it only employs the volume of pulse. Blood pressure, PTT and RAT were acquired from 15 healthy subjects. Each subjects were performed forty trials of each measurement. As a result of those trials, the mean error between oscillometric and RAT measurements for systolic blood pressure was 4.55 pm5.64mmHg. This result showed quite equal with the mean error between oscillometric and PPT measurf:ments, 4.22 pm5.30mmHg, However, it was not obtained a satisfactory result in the relativity of oscillometric to both RAT and PPT measurements for diastolic blood pressure because of personal difference. To conclude, the method of systolic blood pressure estimation noninvasively and unrestrainedly using by RAT may be used as the method by PTT. Nevertheless, additional studies would be necessary for the RAT/PTT estimation of diastolic blood Pressure measurement.